Please use this identifier to cite or link to this item: http://hdl.handle.net/123456789/23727
Title: To study the efficiency of cr-39 based dosimeters
Authors: Ahmad, Iftikhar
Keywords: Chemistry
Issue Date: 2000
Publisher: Quaid I Azam University
Abstract: In the field of solid state nuclear track detectors (SSNTDs), plastic CR-39 is the highly sensitive track recording material .The studies were carried out to see the effect of concentration of etchant (in the range of 4 to 12 M), different etchant and temperature (in the range 50-82°C) on the bulk etch rate ofCR-39. The reported method for bulk etch rate determined (Henke Benton's method) was modified to remove the absorbed etchant and get better results with high accuracy and reduce the systematic error in bulk etch rate. It was also observed that the saturated etchant decreased the bulk etch rate. Whereas the water absorbed before exposure the detector enhances the bulk etch rate. It was also found that the etchant concentration strongly effected on the alpha track diameters. Bulk etch rate of CR-39 depends upon the absorbed dose by y-rays.
URI: http://hdl.handle.net/123456789/23727
Appears in Collections:M.Phil

Files in This Item:
File Description SizeFormat 
CHE 472.pdfCHE 47210.45 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.